Simulation and Measurement of A Small Cosmic Ray Device
-
WANG Rong,
-
HAN Chengdong,
-
ZHANG Yapeng,
-
ZHAO Fengyi,
-
LI Yao,
-
FU Qiang,
-
YIN Jun,
-
NI Fafu,
-
WANG Yanyu,
-
ZHANG Pengming,
-
CHEN Xurong
-
Graphical Abstract
-
Abstract
A small cosmic ray device is introduced in this paper.It has the demonstration function for popularization of science,and can be used mainly to display the μ counts in a real-time measurement,and to monitor the secondary cosmic ray flux at different angles in a very long time scale (years).We briefly show the hardware of the device and the detecting efficiency calculation by simulation.The detecting efficiency for one detector of the device is 93.1%,and the detecting efficiency is 86.6% for the coincidence measurement of two detectors.Based on the count rate by the detector and the simulated efficiency,the secondary cosmic ray flux perpendicular to the ground surface is measured,which is J=29±3 m-2sr-1 s-1.Moreover,with an application of the device,we measured the angular distribution of the secondary cosmic ray rate in Lanzhou City.The resulting angular distribution agrees well with the empirical formula as I(θ)=IH+I0 cosα θ,in which the parameter for the angle-dependence is α=2.42±0.53.
-
-