Study on Crystal Size and Detection Efficiency of HPGe Detector

  • The accurate shape of HPGe detector is needed in order to calculate its detection efficiency with Monte Carlo methods. In our calculation, the influence of the HPGe crystal size on efficiency has been investigated; the final crystal sizes were determined by comparison with experiments and were validated by the experimental efficiency obtained at several source-to-detector positions. The results show that when the crystal dimensions determined are used to calculate the detection efficiency in the energy range 122s1 332 keV, the relative deviations between simulated and most experimental results were within 5%. In addition, the relationship between detection efficiency and axial distance to detector was established.