Advanced Search
LI Wen-jian, BAI Yu-shu, YANG Ying-jie, YUAN Jian-lei, SU Xu, DANG Bing-rong, MA Qiu-feng, ZHOU Li-bin. Dosimetry of 12C Ion Radiation[J]. Nuclear Physics Review, 2006, 23(2): 254-257. DOI: 10.11804/NuclPhysRev.23.02.254
Citation: LI Wen-jian, BAI Yu-shu, YANG Ying-jie, YUAN Jian-lei, SU Xu, DANG Bing-rong, MA Qiu-feng, ZHOU Li-bin. Dosimetry of 12C Ion Radiation[J]. Nuclear Physics Review, 2006, 23(2): 254-257. DOI: 10.11804/NuclPhysRev.23.02.254

Dosimetry of 12C Ion Radiation

  • Dosimetric characteristics of L-α-alanine dosemeter used for dosimetry of ^12C ion radiation have been studied. The experimental results indicate that the alanine dosemeter can be used to measure the ^12C ion radiation. In addition, dose effects of chromosome aberration dicentrics and cenric rings were studied after human peripheral blood being irradiated by ^12C ions; the best regression equation, Y = 0. 858 503D + 0. 361 5 ×10^-2D^2, was obtained within 8.0 Gy.
  • loading

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return